EMC China Lab

Medical Device Muscle Implantation Test Laboratory

Views :
Update time : 2024-06-08

1. Scope of application and principle

This experimental method is to insert the implant into the muscle of the test animal and compare the biological response of the test sample implant with that of the control sample implant to evaluate the biological response of muscle tissue to the implanted material. The control material is the material used in the medical device for which clinical acceptability and biocompatibility have been established.

Medical Device Muscle Implantation Test Laboratory(图1)


2. Muscle Implantation Test standards

Domestic: GB/T 16886.6 Biological evaluation of medical devices Part 6: Local reaction test after implantation;

国外:ISO 10993-6 Biological evaluation of medical devices—Part 6: Tests for local effects after implantation。

3. Sample preparation

The routine preparation of test and control samples is as specified in 3.1. The size of the implant is determined by the size of the muscle group selected. When using the rabbit paraspinal muscles for testing, an implant with a width of 1 to 3 mm and a length of approximately 10 mm is used. The sample should be made with rounded edges and smooth spherical surfaces at both ends. Alternatively, a larger sample with a diameter of 10 mm and a thickness of 3 mm can be surgically implanted. The sample should be made with rounded edges and blunt corners at both ends.

4. Selection of control samples

The control sample can be a marketed control, or a negative control material can be selected according to Appendix A of GB/T 16886.12, such as PE, silicone resin, alumina, and stainless steel.

5. Experimental animals and implantation sites

When implanting the implant into the muscle tissue of rabbits or other animals, the size of the implant site should meet the requirements of the implant. Only one animal can be used for each test. Rabbit paraspinal muscle is the preferred implant site. Smaller samples can also be used in rat gluteal muscle or rabbit femoral muscle.

6. Number of experimental animals used

At least three animals were used per implantation period, with 10 test samples and 10 control samples implanted at adequate implantation sites. The test and control samples used for evaluation should have come from at least three different animals.

If it is desired that the control material produce a response greater than the minimal response, an additional control material known to elicit a minimal tissue response may be implanted in the area contralateral to the test material.

7. Implantation steps

When using the paraspinal muscles of rabbits, implant 4 or 3 test material samples into the muscle on one side of the spine of each rabbit, parallel to the spine, 25-50 mm from the midline, with each implant approximately 25 mm apart. Implant 4 or 3 control material samples on the other side of the spine in the same manner.

8. Implantation cycle

The test period should be determined based on the possible clinical contact time, or the time until the corresponding biological response reaches or exceeds a certain stable state. The selected time point should be explained and demonstrated. Ensure that the biological tissue response reaches a stable state. Short-term selections generally include 1 week and 4 weeks. Long-term selections generally include 13 weeks and 26 weeks. Table 1 shows the standard recommended long-term implantation observation period selection.

Table 1. Selection of observation period for long-term implantation in subcutaneous tissue, muscle, and bone

Animal species

                       Implantation period (weeks)

13

26

52

78

104

Mouse

×

×

×

Rat

×

×

×

guinea pig

×

×

×

Rabbit

×

×

×

×


dog

×

×

×

×

×

sheep

×

×

×

×

×

goat

×

×

×

×

×

pig

×

×

×

×

×


9. Sampling and pathological section preparation

At the end of the experiment, the animals were euthanized and samples were collected after dissection. Pathological sections were made and HE staining was performed.

10. Biological response evaluation

Biological reactions are evaluated by recording macroscopic and histopathological reactions observed at different time points. Compare the reactions of the test sample, control sample or sham operation site. Score the local biological reactions after implantation according to Appendix E of GB/T 16886.6. Determine the tissue reaction of the device according to the table below, compared with the marketed control or the information control.

Table 6. Histopathological results

total score

Organizational response

0.0~2.9

No stimulation

3.0~8.9

Mild irritation

9.0~15.0

Moderate irritation

>15.0

Severe stimulation



Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
What are the CB Certification Body? What are the CB Certification Body?
09 .29.2024
CB Certification is an international system under IECEE for testing electrotechnical product safety,...
EN50689:2021 Testing Services Laboratory EN50689:2021 Testing Services Laboratory
09 .29.2024
EN50689:2021 ensures consumer laser product safety by regulating design, testing, and labeling, cove...
Brazil ANATEL Mobile Phone Charger Testing Standar Brazil ANATEL Mobile Phone Charger Testing Standar
09 .29.2024
ANATEL Act 5155, effective October 2024, updates mobile charger testing with stricter ESD, improved ...
The Difference Between CPC and GCC Certificates The Difference Between CPC and GCC Certificates
09 .29.2024
CPC is for children‘s products (12 and under), requiring CPSC-approved lab testing. GCC covers gener...
Medical Endoscope FDA Certification ISO 10993 Test Medical Endoscope FDA Certification ISO 10993 Test
09 .28.2024
China JJR Laboratory offers FDA certification and ISO 10993 testing for medical endoscopes, ensuring...
Wireless Charger FCC ID Certification Wireless Charger FCC ID Certification
09 .28.2024
Wireless chargers require FCC ID certification, adhering to standards like FCC Part 15 and Part 18, ...
What is USP Class VI for Raw Materials? What is USP Class VI for Raw Materials?
09 .28.2024
USP Class VI ensures biocompatibility for medical plastics by passing stringent tests for low toxici...
FCC ID Certification for Wireless Cameras FCC ID Certification for Wireless Cameras
09 .28.2024
Wireless cameras need FCC ID certification for U.S. sales, adhering to standards like FCC Part 15 an...

Leave Your Message