◆ CISPR 11/EN 55011/GB 4824:
"Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement"
◆ IEC 60601-1-2/YY 0505:
"Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests"
◆ IEC 61326-1/GB T18268.1:
"Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 1: General Requirements"
◆ IEC 61326-2-6/GB T18268.26:
"Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 26: Specific Requirements - In Vitro Diagnostic (IVD) Equipment"
◆ IEC 61000-3-2/GB 17625.1:
"Electromagnetic Compatibility - Limits - Limits for Harmonic Current Emissions (Equipment Input Current ≤16A per Phase)"
◆ IEC 61000-3-3/GB 17625.2:
"Electromagnetic Compatibility - Limits - Limitation of Voltage Changes, Voltage Fluctuations, and Flicker in Public Low-Voltage Supply Systems, for Equipment with Rated Current ≤16A and Subject to Conditional Connection"
◆ Conducted Disturbance Voltage, 150kHz-30MHz, Quasi-Peak/Average
◆ Radiated Disturbance (Magnetic Field), 9kHz-30MHz, Quasi-Peak
◆ Radiated Disturbance (Electric Field), 150kHz-1GHz, Quasi-Peak; 1GHz-18GHz, Peak/Weighted
◆ Harmonic Current Emission Test, According to IEC 61000-3-2
◆ Voltage Fluctuation and Flicker Test, According to IEC 61000-3-3
◆ Electrostatic Discharge Immunity, Contact Discharge 6kV, Air Discharge 8kV
◆ Radiated Radio-Frequency Electromagnetic Field Immunity, Maximum Test Field Strength 10V/m
◆ Electrical Fast Transient/Burst Immunity, Maximum Test Level 2kV
◆ Surge Immunity, Maximum Test Level 2kV
◆ Radio-Frequency Conducted Immunity, Maximum Test Level 10V
◆ Power Frequency Magnetic Field Immunity, Maximum Test Field Strength 3A/m
◆ Voltage Dips, Short Interruptions, and Voltage Variations Immunity, According to IEC 61000-4-11