EMC China Lab

Medical Device EMC Test Items & Standards

Views :
Update time : 2024-04-26

Main Reference Standards

◆ CISPR 11/EN 55011/GB 4824:

   "Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement"

◆ IEC 60601-1-2/YY 0505:

   "Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests"

◆ IEC 61326-1/GB T18268.1:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 1: General Requirements"

◆ IEC 61326-2-6/GB T18268.26:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 26: Specific Requirements - In Vitro Diagnostic (IVD) Equipment"

◆ IEC 61000-3-2/GB 17625.1:

   "Electromagnetic Compatibility - Limits - Limits for Harmonic Current Emissions (Equipment Input Current ≤16A per Phase)"

◆ IEC 61000-3-3/GB 17625.2:

   "Electromagnetic Compatibility - Limits - Limitation of Voltage Changes, Voltage Fluctuations, and Flicker in Public Low-Voltage Supply Systems, for Equipment with Rated Current ≤16A and Subject to Conditional Connection"

 

EMI Test Items

◆ Conducted Disturbance Voltage, 150kHz-30MHz, Quasi-Peak/Average

◆ Radiated Disturbance (Magnetic Field), 9kHz-30MHz, Quasi-Peak

◆ Radiated Disturbance (Electric Field), 150kHz-1GHz, Quasi-Peak; 1GHz-18GHz, Peak/Weighted

◆ Harmonic Current Emission Test, According to IEC 61000-3-2

◆ Voltage Fluctuation and Flicker Test, According to IEC 61000-3-3

 

EMS Test Items

◆ Electrostatic Discharge Immunity, Contact Discharge 6kV, Air Discharge 8kV

◆ Radiated Radio-Frequency Electromagnetic Field Immunity, Maximum Test Field Strength 10V/m

◆ Electrical Fast Transient/Burst Immunity, Maximum Test Level 2kV

◆ Surge Immunity, Maximum Test Level 2kV

◆ Radio-Frequency Conducted Immunity, Maximum Test Level 10V

◆ Power Frequency Magnetic Field Immunity, Maximum Test Field Strength 3A/m

◆ Voltage Dips, Short Interruptions, and Voltage Variations Immunity, According to IEC 61000-4-11


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
Air Cooler CE Certification Testing Air Cooler CE Certification Testing
03 .31.2025
Air coolers require CE certification for Europe. Testing by China JJR Laboratory ensures compliance ...
Electric Hammer CE Certification Testing Electric Hammer CE Certification Testing
03 .31.2025
Electric hammer CE certification testing by China JJR Laboratory ensures compliance with EU standard...
Hair Clipper CE Certification Testing Hair Clipper CE Certification Testing
03 .31.2025
Hair clippers require CE certification for Europe, complying with Low Voltage 2014/35/EU, EMC 2014/3...
Ultrasonic Fishing Machine CE Certification Test Ultrasonic Fishing Machine CE Certification Test
03 .31.2025
The ultrasonic fishing machine, CE certified, follows EU directives and standards. Testing is provid...
Power Adapter Certifications for Thailand Power Adapter Certifications for Thailand
03 .31.2025
Power adapters exported to Thailand require TISI certification. China JJR Laboratory offers testing ...
US Clothing GCC Certificate 16CFR 1610 Standard Te US Clothing GCC Certificate 16CFR 1610 Standard Te
03 .31.2025
US Clothing GCC Certificate 16CFR 1610 flammability test ensures compliance with US safety standards...
US Fire Mask ASTM E2952-23 Standard Test US Fire Mask ASTM E2952-23 Standard Test
03 .31.2025
The Fire Mask ASTM E2952-23 standard tests air-purifying smoke protection devices, ensuring 15 minut...
Teething Toy CPC Certification ASTM F963-23 Testin Teething Toy CPC Certification ASTM F963-23 Testin
03 .31.2025
JJR Lab offers ASTM F963-23 testing for teething toys, covering updates on heavy metals, phthalates,...

Leave Your Message