EMC China Lab

ISO 18562 Biocompatibility Assessment

Views :
Update time : 2024-09-24

Like all medical devices, those that come into contact with respiratory gases require a biocompatibility assessment. While the ISO 10993 series is well-known in the medical device industry, the iso 18562 series, specifically for respiratory gas pathways, remains unfamiliar to many manufacturers despite its existence since 2017.

 

Structure of ISO 18562 Series

The starting point for biocompatibility assessment is ISO 18562-1, which outlines the testing process, planning, and data analysis. The tests themselves reference:

- ISO 18562-2: Particulate Matter Emission Testing

- ISO 18562-3: Volatile Organic Compounds (VOC) Emission Testing

- ISO 18562-4: Testing for Leachables in Condensate

 

Part 1: Evaluation and Testing Within a Risk Management Process

The simplified process is as follows:

- Assess the components and materials in contact with respiratory gases for relevant biocompatibility data.

- Determine the required tests, including the test devices.

- Conduct the tests.

- After obtaining emission data, calculate the dosage for patients and compare the risks with the clinical benefits of the medical device.

 

When selecting tests, consider not only those in ISO 18562 Parts 2 to 4 but also other substances that may raise concerns, which regulatory authorities may require to be evaluated:

- Semi-volatile organic compounds (SVOC) and very volatile organic compounds (VVOC)

- Ozone generated during operation of electrostatic or electromechanical components

- CO and CO2

- Leachables, including not only condensate but also anesthetics or other substances delivered via the respiratory tract, such as inhaled medications.

 

Manufacturers should determine the usage time for the test setup, influencing both the testing timeline and providing input for risk assessment.

 

Part 2: Tests for Emissions of Particulate Matter

ISO 18562-2 specifies methods for collecting particulate matter added to respiratory gases, including:

- A filtration method ensuring input air is clean, collecting particulate matter through the filter and measuring emissions by weighing the filter.

- A two-filter method when clean input air cannot be ensured, measuring emissions by weighing both filters before and after testing.

 

Additionally, particle counters can replace filters. The standard sets maximum values as follows:

- For particles with a diameter ≤ 2.5 micrometers, > 12 micrograms/m³;

- For particles with a diameter ≤ 10 micrometers, > 150 micrograms/m³.

Important Note: Devices should operate at maximum clinically relevant flow rates.

 

Part 3: Tests for Emissions of Volatile Organic Compounds (VOCs)

ISO 18562-3 outlines methods for collecting VOCs and establishes TTC limits based on exposure time. The first 24 hours allow 360 µg/d, followed by 120 µg/d for 29 days, and 40 µg/d after 30 days.

 

For measuring VOCs, ISO 18562-3 recommends collecting the following data:

- For continuous flow devices, collect at the stated breathing volume.

- For intermittent flow devices, collect in a clinically relevant manner.

 

Setting the device at the highest rated ambient temperature, gas outputs should be sampled at different times as per the testing plan, followed by analysis to determine total VOC emissions and individual VOC levels.

 

Part 4: Tests for Leachables in Condensate

Devices that allow condensate to enter patients must also undergo leachables testing. Collect leachables using one of the following sampling methods:

- Generating and collecting condensate under clinically relevant conditions.

- Surface circulating water at representative clinical temperatures.

 

Extract internal gas contact surfaces per ISO 10993-12. Due to the difficulty of the first two methods, most manufacturers use the extraction method from iso 10993-12. After obtaining extracts, perform four tests:

- Determine metal ion content.

- Conduct cytotoxicity testing per iso 10993-5.

- Conduct sensitization testing per iso 10993-10.

- Identify and quantify organic impurities.

 

Toxicological Risk Assessment

The final step is returning to ISO 18562-1 for a toxicological risk assessment. This assessment includes all test results and historical data, calculating the doses patients will receive. If a device applies to multiple patient categories, calculate doses for all. Once patient doses are known, compare them with existing or newly generated maximum exposure data to determine the risks posed by the medical device. This provides another input for the risk-benefit analysis for each medical device.

 

Is ISO 10993 Irrelevant to Respiratory Gas Pathway Devices?

No, only respiratory gases require assessment per ISO 18562; all parts of the device that may contact patients (or users or bystanders) must also be evaluated per ISO 10993.


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
Which Laboratory Can Process Lithium Battery EN 50 Which Laboratory Can Process Lithium Battery EN 50
07 .22.2026
Need EN 50604-1 certification for LEV lithium batteries? Discover the mandatory safety testing requi...
Bicycles on Amazon: 16 CFR 1512 and GCC Certificat Bicycles on Amazon: 16 CFR 1512 and GCC Certificat
07 .22.2026
Selling bicycles on Amazon US requires 16 CFR 1512 compliance and a GCC certificate. Learn about man...
What is the Cycle for UL 2271 Certification for Li What is the Cycle for UL 2271 Certification for Li
07 .22.2026
Discover UL 2271 certification requirements for US Amazon LEV lithium batteries. Explore electrical,...
The Two Certification Schemes of Hong Kong OFTA The Two Certification Schemes of Hong Kong OFTA
07 .22.2026
Navigate Hong Kong OFTA certification for telecom and radio equipment. Ensure market access complian...
What is the difference between ANSI C18.3M and UL4 What is the difference between ANSI C18.3M and UL4
07 .22.2026
Amazon sellers of button and coin batteries must comply with Reese‘s Law. Submit ISO 17025 reports p...
Amazon Button and Coin Cell Battery ANSI C18.3M Ce Amazon Button and Coin Cell Battery ANSI C18.3M Ce
07 .22.2026
Ensure Amazon compliance for button & coin cell batteries with mandatory ANSI C18.3M & UL 42...
Button Cell Battery ANSI C18.3M Testing Button Cell Battery ANSI C18.3M Testing
07 .22.2026
JJR LAB offers comprehensive ANSI C18.3M testing for button cell lithium batteries, including electr...
Lithium Battery IEC 62133-2:2017 Certification Gui Lithium Battery IEC 62133-2:2017 Certification Gui
07 .21.2026
Ensure global compliance for portable lithium batteries. JJR Lab provides expert IEC 62133-2:2017 te...

Leave Your Message