EMC China Lab

Hemocompatibility Testing ISO 10993

Views :
Update time : 2024-11-28

ISO 10993-4:2017 Interpretation of Hemocompatibility Testing Standards

Medical devices that come into direct or indirect contact with human circulating blood must undergo hemocompatibility testing if there is insufficient data to demonstrate compliance with blood compatibility requirements. ISO 10993-4:2017 is a key reference standard for such tests. Today, China's JJR Laboratory will interpret the requirements of this standard and explore the testing protocols.

 

Overview of Hemocompatibility Testing

Hemocompatibility testing evaluates the effects of medical devices or materials in contact with blood or blood products. When selecting and designing test protocols, factors such as product design, clinical use, environment, and risk-benefit analysis should be considered.

 

ISO 10993-4:2017, as an essential reference, primarily includes three components:

1. Classification of blood-contacting medical devices.

2. Fundamental characteristics for assessing blood responses.

3. Selection of test methods.

 

Classification of Blood-Contacting Devices

Non-blood-contact devices refer to those that do not directly or indirectly interact with blood that remains in or is returned to the body, such as in vitro diagnostic reagents and blood collection tubes.

 

External communicating devices generally interact with circulating blood and connect to the human vascular system. Examples include blood collection devices, catheters, blood storage or transport equipment like blood bags and transfusion tubing, vascular catheters, and intravenous extensions. Other examples include atherectomy devices, blood monitoring systems, cardiopulmonary bypass circuits, therapeutic blood separation equipment, extracorporeal membrane oxygenators, dialysis devices, cardiac interventional devices, and vascular guidewires.

 

Implantable devices involve prolonged contact with blood and include annuloplasty rings, arteriovenous shunts, implantable monitors, ventricular assist devices, artificial hearts, embolization devices, synthetic vascular grafts, implantable defibrillators, pacemaker leads, stents, and similar products.

 

Blood Reaction Characteristics

After identifying the device category, it is essential to understand the blood reaction characteristics of the product to select appropriate test methods. Two critical points should be noted:

- The standard provides a decision flowchart to determine whether hemocompatibility testing is necessary.

- The standard lists device examples with corresponding test categories.

 

Answering the following four questions can help determine whether hemocompatibility testing is required:

1. Does the device directly or indirectly contact circulating blood? If not, this standard does not apply; if yes, proceed to the next question.

2. Do the device materials match those of marketed products in terms of processing, structure, physicochemical properties, sterilization methods, and clinical usage? If all match, no testing is needed; otherwise, proceed further.

3. Can differences be evaluated and found acceptable? If yes, testing is unnecessary; if not, further risk evaluation is needed.

4. Is the risk assessment data on blood compatibility acceptable? If yes, testing is unnecessary; otherwise, testing is required according to relevant guidelines.

 Hemocompatibility Testing ISO 10993(图1)

Examples of devices requiring testing include blood monitoring devices, which may require hemolysis, coagulation, platelet activation, and hematology tests.

 Hemocompatibility Testing ISO 10993(图2)Hemocompatibility Testing ISO 10993(图3)

Parameters to consider include the structural design, physicochemical properties, surface characteristics, sterilization conditions, duration of contact, temperature, and blood flow or anticoagulation conditions. For implantable products, testing should closely simulate clinical conditions.

 

Description of Testing Methods

Generally, hemocompatibility testing includes two main categories:

1. Hemolysis Tests: These may be material-mediated or mechanically mediated.

2. Thrombosis Tests: These include in vitro tests for coagulation, platelet activation, and complement activation, as well as in vivo and ex vivo tests.

 

Key considerations for in vitro tests include hematocrit, anticoagulant type and dosage, blood sample preparation, temperature, pH, flow conditions, and blood/surface area ratios.

 Hemocompatibility Testing ISO 10993(图4)

Ex vivo tests are essential for extracorporeal devices like dialysis systems. These tests can monitor platelet adhesion, thrombus formation, and fibrinogen deposition. Real-time blood flow changes can indicate thrombus deposition and embolization.

 

For implantable devices like vascular stents, in vivo animal studies should simulate clinical conditions. Key evaluation criteria include device patency and histopathological assessment after removal. Advanced techniques such as angiography or intravascular ultrasound (IVUS) may also be used.

 

Conclusion

ISO 10993-4:2017 outlines the classification of blood-contacting medical devices, fundamental blood reaction characteristics, and standard testing methods. However, detailed guidance on test designs, such as in vivo thrombosis or platelet activation assays, requires consulting Annexes A-G of the standard.

 

References

[1] ISO 10993-4:2017 Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with blood


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
Carbon Monoxide Alarm Amazon Test Report Carbon Monoxide Alarm Amazon Test Report
02 .20.2025
JJR Laboratory provides testing for carbon monoxide alarms to meet Amazon’s safety standards, ensuri...
ASTM F2057 Standard ISO 17025 Test Report ASTM F2057 Standard ISO 17025 Test Report
02 .20.2025
ASTM F2057-2023 ensures clothing storage unit safety by preventing tip-overs, requiring stability, f...
How to get a Lighting UL Test Report? How to get a Lighting UL Test Report?
02 .20.2025
Amazon requires a UL test report for lighting products. JJR Laboratory provides testing per UL stand...
What is the UL 60950 Test Report? What is the UL 60950 Test Report?
02 .20.2025
UL 60950 is a North American safety standard for IT power supplies, covering insulation, voltage, an...
UL 8800 Certification Testing Requirements UL 8800 Certification Testing Requirements
02 .20.2025
UL 8800 Testing by JJR Lab: The updated UL 8800 standard improves safety for horticultural lighting ...
How to get the UL8800 Test Report for Lighting Fix How to get the UL8800 Test Report for Lighting Fix
02 .20.2025
To obtain the UL8800 test report for plant lights, submit your product to JJR Laboratory for testing...
What is UL1647 Certification? What is UL1647 Certification?
02 .20.2025
UL1647 certification ensures the safety of electric massagers and fitness equipment. JJR Lab provide...
UL1647 Compliance Certification for Massage Device UL1647 Compliance Certification for Massage Device
02 .20.2025
JJR Laboratory offers UL1647 testing for massage devices to meet Amazon‘s compliance requirements. W...

Leave Your Message