We understand that you are looking for an EMC (Electro Magnetic Compatibility) testing laboratory in Canada. However, we still want to help you understand us; because compared to local Australian laboratories, we can save you 30% on testing and certification costs.
JJRLAB is a Chinese EMC electromagnetic compatibility testing laboratory; we are an ISO/IEC 17025 and GLP laboratory. Welcome to learn more about us.
Radiated Emissions
(Radiated Emission )
Applicable standards
C1SPR1J、CISPR 14-1 x CISPR15、CISPR 16、CISPR32、IEC/EN 60601-1-2, IEC/EN 61000-6-3/-4 、fcc part 15x ICES-003、EN 310489、EN 300386、EN 55011 、EN 55014-1、EN 55015、EN 55032、GB/T 9254^ GB/T 13837、GB/T 17743、GB/T 17799.3/.4、GB 4343.1、GB4824、YY 0505
Measuring range
30MHz-40GHz
Main technical requirements of the site
Normalized site attenuation NSAW≤±4dB, site voltage standing wave ratio SVSWRW≤±6dB
Antenna height 1m-4m, turntable angle 0-360 degrees, ambient level at least 6dB lower than the limit
Shielding Effectiveness SE:
0.014MHz-1 MHz >60dB
1MHz-1000Hz>90dB
1GHz-18GHz >80dB
Radiated immunity
(Radiated Immunity)
Applicable standards
IEC/EN 61000-4-3
GB/T 17626.3
Measuring range
80MHz-6GHz 10V/m
Main technical requirements of the site
At least 75% of the 16 points of site uniformity are between 0-6dB
Conducted emissions
(Con ducted Emission)
Applicable standards
CISPR 11, CISPR 14-1、CISPR 15、CISPR 16、CISPR 32、IEC/EN 60601-1-2. IEC/EN 61000-6-3/-4、FCC PART 15x ICES-003、EN 310489、EN 300386、EN 55011、EN 55014-1、EN 55015、EN 55032、GB/T 9254、GB/T 13837、GBAT 17743、GB/T 17799.3/.4、GB 4343.1、GB 4824、YY 0505
Measuring range
150kHz-30MHz (power supply side, power borrowing side)
Main technical requirements of the site
LISN grounding resistance is less than 2.5mΩ
The ambient level is at least 6dB below the limit
Harmonic Current Emissions
Voltage Fluctuations and Flicker
Applicable standards
IEC/EN 61000-3-2、GB 17625.1
IEC/EN 61000-3-3、GB 17625.2
Measuring range
1-40th harmonic
Pst-10Min, Plt-2H
Main technical requirements of the site
Test voltage within ±2%
The total harmonic distortion of the power supply voltage is less than 3%
The current accuracy is above ±1% and the measured phase angle error does not exceed ±2.
Electrostatic discharge immunity
(Electrostatic Discharge Immunity )
Applicable standards
IEC/EN 61000-4-2、ISO 10605
GB/T 17626.2、GB/T 19951
AEC-Q100-002-Rev.D Jul.2003 ESDA ANSI/EOS/ESD-SIM5.1 -2001 IEC61340-3-1 Ed1.0 2002、IEC 60749-26 Ed1.0 2003、JEDEC JESD22-A114C Jan.2005 、JEITA EIAJ ED-4701/300 Aug.2001、JEDEC JESD22-A115A
Measuring range
Finished product ±30KV
Module ±8KV
Main technical requirements of the site
470KΩ resistors at both ends of the coupling plate and the ground reference plane
Temperature: 15℃-35℃: humidity 30%RH-60%RH, atmospheric pressure 86kPa-106kPa
Voltage polarity: positive, negative
Duration: no less than 5S
Discharge mode: single, continuous
Electrical fast pulse group interference
(Electrical Fast Transients Immunity )
Applicable standards
IEC/EN 61000-4-4
GB/T 17626.4
Measuring range
±5.5KV (power supply, signal end)
Main technical requirements of the site
Polarity: positive, negative
Output type: coaxial output, 50Ω
DC blocking capacitor: 10 (1 ± 20%) nF
Repetition frequency: 5&100 (1±20%) kHz
Pulse group duration: 15 (1 ± 20%) ms at 5kHz, 0.75 (1 ± 20%) ms at 100kHz
Pulse group period: 300 (1 ± 20%) ms
Lightning surge immunity
(Surge Immunity)
Applicable standards
IEC/EN 61000-4-5
GB/T 17626.5
Measuring range
±6.6KV (power supply end, multiple end)
Main technical requirements of the site
Polarity: positive, negative
Wavefront time
Open circuit: 10us±30%, 1.2us±30%
Short circuit: 8us±20%, 5us±20%
Half peak time
Open circuit: 700us±20%, 50us±20%
Short circuit: 20us±20%, 320us±20%
Phase angle: 0° -360°
Conducted Immunity
(Con ducted Immunity )
Applicable standards
IEC/EN 61000-4-6
GB/T 17626.6
Measuring range
150KHz-230MHz, 10V (power supply, signal)
Main technical requirements of the site
Polarity: positive, negative
Wavefront time
Open circuit: 10us ±30%, 1.2us±30%
Short circuit: 8US±20%, 5US±20%
Half peak time
Open circuit: 700us±20%, 50us±20%
Short circuit: 20us±20%, 320us±20%
Phase angle: 0° -360°
Power frequency magnetic field immunity
(Power Frequency Magnetic Field Immunity )
Applicable standards
IEC/EN 61000-4-8
GB/T 17626.8
Measuring range
Short-term 1000A/m, continuous 100A/m
Main technical requirements of the site
The EUT should be placed within a magnetic field uniformity space of ±3dB.
Continuous mode: 1A/m~100A/m, 3S-99999S
Short-time mode: 101A/m~1000A/m, 1s-3s
Additional current waveform distortion rate: <8%
Current waveform: 50Hz or 60Hz sine wave
Voltage drop, instantaneous interruption, and interference immunity
(Voltage Dips and Short Interruptions Immunity )
Applicable standards
IEC/EN 61000-4-11
GB/T 17626.11
Measuring range
0-100%
Main technical requirements of the site
The frequency of the power supply test voltage is within ±2% of the rated frequency.
When the generator carries 100 load, the actual voltage instantaneous peak overshoot/undershoot is less than 5%UT
When the generator carries 100 load, the voltage rise and fall time during the mutation process is 1us~5us
Phase change 0-360
The phase relationship between voltage sag and interruption and power frequency is less than ±100
Zero crossing control of generator ±100
DC voltage drop instantaneous interruption immunity
(DC-Voltage Dips and Short Interruptions Immunity )
Applicable standards
IEC/EN 61000-4-29
GB/T 17626.29
Measuring range
0-100%
Main technical requirements of the site
Output voltage range (Uo) ≤ 360V
Output voltage changes with load <5%
Ripple content < 1% of output voltage
When the generator is loaded with 100Ω, the overshoot/undershoot of the output voltage is less than 10% of the voltage change.
When the generator carries a 100Ω load, the voltage rise and fall time is 1us~50us
Output current (steady state) up to 25A