EMC China Lab

China Mask Certification Testing Laboratory

Views :
Update time : 2025-03-28

In response to market and consumer concerns about mask quality and safety, as well as the demand for mask exports overseas, China JJR Laboratory has invested over one million USD in advanced mask testing equipment and actively sought recognition from national and international organizations to provide fast, flexible, and professional internationally recognized mask testing services. Today, JJR has obtained dual recognition from the China National Accreditation Service for Conformity Assessment (CNAS) and the International Accreditation Service (IAS), offering mask testing services accredited by the International Laboratory Accreditation Cooperation (ILAC).

 

China JJR Laboratory provides comprehensive international standard testing such as: Australia AS4381: 2015, China YY0469-2011, YY/T0969-2013, and GB15979, European Union EN 14683: 2019+AC: 2019, and the United States ASTM F2100-19, meeting the requirements for mask exports to different countries and registration.

 

ASTM F2100-19 Testing Methods:

- Bacterial Filtration Efficiency (BFE) %:

- Level 1: ≥95

- Level 2: ≥98

- Level 3: ≥98

- Testing Method: ASTM F2101

 

- Particulate Filtration Efficiency (PFE) % @0.1 micron:

- Level 1: ≥95

- Level 2: ≥98

- Level 3: ≥98

- Testing Method: ASTM F2299

 

- Pressure Differential (mmH₂O/cm²):

- Level 1: <5.0

- Level 2: <6.0

- Level 3: <6.0

- Testing Method: EN 14683

 

- Synthetic Blood Penetration Resistance (mmHg):

- Level 1: 80

- Level 2: 120

- Level 3: 160

- Testing Method: ASTM F1862

 

- Flammability Test:

- Level 1: Class 1

- Level 2: Class 1

- Level 3: Class 1

- Testing Method: 16 CFR 1610

 

EN 14683: 2019 Testing Methods:

- Bacterial Filtration Efficiency (BFE) %:

- Level I: ≥95

- Level II: ≥98

- Level IIR: ≥98

- Testing Method: EN 14683

 

- Pressure Differential (Delta P) (Pa/cm²):

- Level I: <40

- Level II: <40

- Level IIR: <60

- Testing Method: EN 14683

 

- Splash Resistance (kPa):

- Level I: Not Required

- Level II: Not Required

- Level IIR: ≥16.0 (120 mmHg)

- Testing Method: EN 14683

 

- Microbial Cleanliness (CFU/g):

- Level I: ≤30

- Level II: ≤30

- Level IIR: ≤30

- Testing Method: EN 14683

 

AS 4381: 2015 Testing Methods:

- Bacterial Filtration Efficiency (BFE) %:

- Level 1: ≥95

- Level 2: ≥98

- Level 3: ≥98

- Testing Method: EN 14683

 

China Mask Certification Testing Laboratory(图1)


- Pressure Differential (mmH₂O/cm²):

- Level 1: <4.0

- Level 2: <5.0

- Level 3: <5.0

- Testing Method: EN 14683

 

- Synthetic Blood Penetration Resistance (mmHg):

- Level 1: 80

- Level 2: 120

- Level 3: 160

- Testing Method: ISO 22609

 

Biocompatibility ISO 10993 Testing Methods:

- Cell Toxicity: ISO 10993-5

- Irritation: iso 10993-10

- Skin Sensitization: ISO 10993-10

 

In addition to the above standards, our center also offers various mask tests to meet different market requirements, including:

 

- Viral Filtration Efficiency (VFE) %

- GB 15979 Microbial Indicator Requirements

- Antibacterial Efficiency Testing

- Antiviral Efficiency Testing

 

Mask Testing Equipment:

To meet the large demand for mask testing and to ensure timely reporting, China JJR Laboratory has invested over one million USD in advanced mask testing equipment to immediately improve the mask testing capacity and speed.

 


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
Amazon UL Standard Test Report Amazon UL Standard Test Report
04 .26.2025
China JJR Lab provides UL testing for Amazon listings, ensuring ISO 17025 compliance for powered pro...
When Can FCC ID Modifications Be Filed? When Can FCC ID Modifications Be Filed?
04 .26.2025
FCC ID modification includes Class I, II, III, and Change ID. JJR Lab China provides reliable, effic...
LoRa Certification Testing Laboratory LoRa Certification Testing Laboratory
04 .26.2025
LoRa Certification Testing by China JJR Lab: We offer full testing for LoRa modules and gateways acr...
Blood Pressure Monitor Certification Testing Servi Blood Pressure Monitor Certification Testing Servi
04 .26.2025
Blood pressure monitor testing per global and China standards, including EMC, safety, performance, a...
ECG Device Certification Testing ECG Device Certification Testing
04 .26.2025
ECG device certification testing, including EMC, safety, performance, biocompatibility, and steriliz...
Pulse Oximeter Certification and Testing Standards Pulse Oximeter Certification and Testing Standards
04 .26.2025
Pulse oximeters measure SpO₂ using light technology. JJR Laboratory in China provides certification ...
IVD Medical Device GB 4793:2024 Test Report IVD Medical Device GB 4793:2024 Test Report